Maximum syringe size:
Although a maximum of a 60 cc syringe is specified, the pump can hold up to a 140 cc syringe but can only open up to about 120 cc lengthwise
High pressure notes:
The NE-1000 produces about 15 lbs. of linear force at top speed and about 35 lbs. at slow speeds. 10 ml/hr is an example of slow speed. The pressure that can be generated is relative to the syringe size. On a 60 ml syringe, this would roughly produce about 30 psi.
The high pressure version, NE-1010, will approximately double the pumping force at the same speeds.
Description:
Holds 1 Syringe up to 60 cc*
Infusion rates from 0.73 µL/hr ( 1 cc syringe) to 2100 ml/hr (60 cc syringe)
Easy-to-use keypad interface
Space Saving Chassis: Foot print size of only 5 3/4” x 8 3/4”
Includes all the advanced functions and programming features of the NE-1000 family
Special Application: Continuous Infusion (Push-Pull) and Dual Syringe Pump
Using 2 NE-1000 pumps attached with a cable, one pump will infuse while the other pump refills or have simultaneous infusion.
Features:
Fully programmable; automation capable
Operates stand-alone or from a computer
Infusion and withdrawal
Set a single pumping rate and/or dispensing volume
Program up to 41 pumping phases that change pumping rates, set dispensing volumes, insert pauses, control and respond to external signals, sound the buzzer
Network, control, and monitor up to 100 pumps with one computer
Worldwide power supplies available
Motor stall detection
Dispensing accuracy of +/-1%
Unlimited lifetime technical support
Two year warranty
120/240V
CE